Technical specifications:
- Mass range: 1-300 amu
- Gas connection: quartz capillary
- Gas inlet: continuously open
Pressure reduction: 2 stages, fractionation-free
Gas flow rate: 1-2 sccm
- Sample pressure: 1000 mbar
- Operating temperature capillary: up to 200 °C
- Analyzer: QMA 200 M
- Detector: C-SEM/Faraday
- Mass spectrometer electronics: QME 200 M
- Software: QuadStarTM
ThermoStarTM for coupling to thermobalance
- reactive and condensable gases are detectable even in small concentrations
- minimal contact of medium with metal parts
- temperature-controlled gas line (quartz capillary)
- no chemical reaction
- certain identification of unknown gases
Principle construction of ThermoStarTM with open gas inlet
- heating for conditioning of analysis
- heated gas inlet chamber
- temperature-controlled test gas line
- PrismaTM quadrupole mass spectrometer
QuadStarTM Analysis Software
- qualitative analysis of all gases evolving during thermal analysis
- simultaneous data acquisition from a thermal analyzer via Analog Input Module (e.g. temperature, sample weight, differential weight loss)
- presentation of mass spectrometric data as a function of the thermobalance temperature
- convenient data evaluation supported by a spectrum library, statistical and magnifying functions
- data exchange with other programs via binary code or ASCII files; QuadStarTM also supports DDE.